Specifications
|
Size Range:
|
0.10μm - 1.0μm, optional ranges up to 5.0μm
|
Channel Sizes:
|
Standard: 0.10, 0.15, 0.2, 0.25, 0.3, 0.5, 0.7, 1.0μm
|
Flow Rate:
|
1.0 CFM (28.3 LPM)
|
Counting Efficiency:
|
50% @ 0.1μm; 100% for particles > 0.15μm (per JIS)
|
Light Source:
|
Laser Diode
|
Zero Count Level:
|
<1 count / 5 minutes
|
Concentration Limits:
|
500,000 Particles / ft3 (17,150,000 / m3) @ 5% Coincidence Loss
|
Calibration:
|
NIST Traceable, per ISO 21501-4 (Unit-to-Unit Accuracy and Repeatability)
|
Count Modes:
|
Automatic, Manual, Beep, Concentration, Cumulative/Differential, Real Time (with LMSNet or LMS Express RT software)
|
Data Storage:
|
3,000 Sample Records, Rotating Buffer (particle and environmental data with location and time)
|
Communication Modes:
|
USB, Ethernet, RS-485 and Secure Data Transfer (LSDT) to and Storage on USB Flash Drive
|
Supporting Software:
|
LMS XChange, LMS Express and Express RT; LMS Standard; LMS Pharma; CREWin
|
4-20mA Analog Sensors:
|
Optional Sensors (up to 4) Include Temperature/Relative Humidity, Air Velocity and Differential Pressure
|
Touch Screen Display:
|
5.7-inch (14.47 cm) Color, 320x240
|
Printer:
|
Optional, Built-in, Thermal
|
Reports:
|
FS-209E, ISO-14644-1 and EU GMP Annex 1
|
Enclosure:
|
Stainless Steel
|
Alarm:
|
Integrated Adjustable Alarm and External Alarm Connector. Alarms on Counts, Low Battery, Sensor Failure
|
Sample Output:
|
Internally Filtered to 100% @ 0.1μm, Optional HEPA Filtration for Cooling System Exhaust
|
Vacuum Source:
|
Internal Pump, Automatic Flow Control
|
Power:
|
100-240 VAC, 50-60 Hz
|
Battery:
|
Optional Internal Li-Ion Removeable/Rechargeable Battery
|
Dimensions:
|
14.65” (l) x 8” (w) x 10.4” (h) [37 x 20 x 26 cm]
|
Weight:
|
With Battery: 26 lbs (11.8 kg) Without Battery: 23 lbs (10.43 kg)
|
Environmental Conditions
|
Operating:
|
50°F to 104°F (10°C to 40°C) / 20% to 95% non-condensing
|
Storage:
|
14°F to 122°F (-10°C to 50°C) / Up to 98% non-condensing
|
Accessories
|
Included:
|
Operating Manual on CD; Isokinetic Sample Probe with Tripod and Tubing; Purge Filter; Power Cable and Spare Fuse; LMS XChange Software; Printer; Printer Paper
|
Optional:
|
Printed Operating Manual; Network Adapter; Wireless Network Adapter; Internal Removeable, Rechargeable Battery; Carrying Case with Wheels; Cleanroom Cart; Temperature/Relative Humidity Probe; Air Velocity Probe; Differential Pressure Probe; Filter Scanning Probe; High Pressure Difuser; 32-Port Universal Manifold; LMS Express and Express RT Software; LMSNet Software; Validation Documentation
|